Product Description
Bante A130 Laboratory pH/Ion Meter Benchtop
High-precision laboratory pH/ion meter with 5 ion concentration measurement methods and the water hardness measurement mode, setup menu contains 18 options, accuracy: 0.002 pH.
pH Mode:
• High-precision laboratory pH/ion meter is equipped with a 7 inch TFT display.
• 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers.
• Solution temperature coefficient compensates for pure water measurements and references
the pH to 25°C.
ORP Mode:
• 1 point offset calibration allows adjusting the displayed value to a known standard.
• Relative and absolute millivolt modes ensure the reliable ORP measurements.
Ion Concentration Mode:
• 2 to 5 points calibration, including the selection of 8 concentration points.
• Electrode management can store and recall up to 3 electrode parameters.
• Selectable ion measurement methods (direct reading, known addition, known subtraction,
sample addition, sample subtraction) and concentration units (ppm, mg/L, mol/L, mmol/L).
Water Hardness Mode:
• 2 to 5 points calibration from low to high concentrations.
• Selectable measurement units (german degree, english degree, french degree, mmol/L and
mg/L) are used for professional water hardness measurements.
Features Bante A130 Laboratory pH/Ion Meter Benchtop :
- Automatic temperature compensation ensures accurate readings over the entire range.
- Auto-read function senses and locks the measurement endpoint.
- Timed interval readings send measurement data to a PC or printer.
- Limit alarm automatically alerts when measurements exceed the specified range.
- Calibration due alarm reminds user to calibrate the meter regularly.
- Calibration log shows the date, time, calibration points, offset and pH slope.
- Password protection prevents unauthorized calibration and settings.
- Expanded memory stores or recalls up to 1000 data sets.
- Reset function automatically restores all settings to the factory defaults.
Specification Bante A130 Laboratory pH/Ion Meter Benchtop :
Model | A130 | |
pH | ||
Range | -2.000 to 20.000 pH | |
Resolution | 0.001 / 0.01 pH, selectable | |
Accuracy | ±0.002 pH | |
Calibration | 1 to 5 points | |
pH Buffer Options | USA / NIST / DIN / 5 custom buffers | |
ORP | ||
Range | -2000.0 to 2000.0 mV | |
Resolution | 0.1 mV | |
Accuracy | ±0.2 mV | |
Calibration | 1 point | |
Measurement Modes | Relative / Absolute mV | |
Ion Concentration | ||
Range | 0.001 to 30000 (depending on the range of ISE) | |
Resolution | 0.001, 0.01, 0.1, 1, automatic | |
Accuracy | ±0.5% F.S. (monovalent), ±1% F.S. (divalent) | |
Calibration | 2 to 5 points | |
Calibration Solutions | 0.001 / 0.01 / 0.1 / 1 / 10 / 100 / 1000 / 10000 | |
Measurement Units | ppm, mg/L, mol/L, mmol/L | |
Measurement Methods | Direct reading, known addition, known subtraction, | |
sample addition, sample subtraction | ||
Electrode Management | 1 to 3 | |
Water Hardness | ||
Range (Concentration) | 0.05 to 200 mmol/L | |
Range (German Degree) | 0 to 1120°dH | |
Range (English Degree) | 0 to 1404°e | |
Range (French Degree) | 0 to 2000°fH | |
Range (Ca²⁺) | 0 to 8000 mg/L | |
Range (CaCO₃) | 0 to 19999 mg/L | |
Range (CaO) | 0 to 11200 mg/L | |
Resolution | 0.001, 0.01, 0.1, 1, automatic | |
Accuracy | ±1% F.S. | |
Calibration | 2 to 5 points | |
Calibration Solutions | 0.01 / 0.1 / 1 / 10 / 100 mmol/L | |
Temperature | ||
Range | 0 to 105°C / 32 to 221°F | |
Resolution | 0.1°C/°F | |
Accuracy | ±0.5°C / ±0.9°F | |
Calibration | 1 point | |
Calibration Range | Reading ±10°C/°F |
Bante A130 Datasheet
CV. SYARIFTAMA GLOBAL INDONESIA adalah distributor dan Supplier Bante A130 di Jawa Barat & Jakarta dan juga menjual Brand Bante dll. lihat produk kami lainnya di sini KMT,SGI,SGM atau hubungi kami lewat Whatsapp 0821-3010-0456. atau via email [email protected], [email protected] & [email protected]
Dalam pengiriman produk yang pelanggan beli, sebelumnya sudah kami cek untuk memastikan produk dalam keadaan baik dan siap kirim.