ORP PH Temperature Meter 3 in 1 ORP 3569

Product Description

ORP PH Temperature Meter 3 in 1 ORP 3569

1. ORP – Potensial Pengurangan Oksidasi
2. PH – Kadar Asam Basa, dan
3. Temperature – Suhu Air

Specifications ORP PH Temperature Meter 3 in 1 ORP 3569 :

Measuring range :
PH: 0.00-14.00pH
ORP:-1999-1999mV
Temp:0-50℃

Resolution :
PH:0.01pH
ORP:1mV
Temp:0.1℃

Acuracy :
PH:±0.1pH
ORP:±5mV
Temp:±1℃

Automatic Temperature Compensation.O℃~50℃
Batery Type:4×1.5V(AG13)
Operating Temperature:0℃~50℃
Calibration:Manual.at one point through trimmer
Dimensions:188mm×35mm×35mm
Weight: 98g

How to Calibrate

PH

  • Rinse the ph electrode first with distilled water,and suck it with filter paper.
  • Turn on the meter. Press the “pH/ORP”button,choice the PH for measuring.
  • Immerse the ph electrode in pH6.86(or pH7.00)buffer solution,gently shake the meter and wait for a stable display.
  • Regulate the trimmer with the screwdriver provided until display shows “6.86”(“7.00”)at buffer solution temperature of 25℃.
  • Rinse the ph electrode with distilled water,and suck it with filter paper (not included)
  • Immerse the pH electrode in pH4.01 or pH9.18(pH10.0)buffer solution,gently shake the electrode and waiting for the reading to stabilize.The showing value and buffer solution’s ph value is within error range.
  • The calibration of the instrumen’s pH range is now complete.

ORP

  • Turn on the meter.Press the”pH/ORP” button,choice the ORP for measuring.
  • Rinse the electrode with distilled water.
  • Immerse the electrode into the fresh 256mV ORP standard solution,then stir it gently,
  • Allow the reading to stabilize and with a small screwdriver turn the calibration trimmer until the display shows”256″.
  • Rinse the electrode with distilled water and suck it with filter paper.
  • Switch off the meter.Replace the battery compartment cap.
  • The calibration of the instrument’s ORP range is now complete.

Important:

  • The ph meter should be recalibrated under following situations;
  • The meter has been used for long time since the last calibration or use
  • The meter has been used in particularly continual conditions.
  • The utmost accuracy is required.

——————————————————————————————————————————————————————————————————

Jual ORP PH Temperature Meter 3 in 1 ORP 3569 harga kompetitif Tentunya Gratis antar untuk Area Jakarta dan dapat dikirim keseluruh Indonesia, Jika Anda butuh bantuan silahkan hubungi Customer Service (026) 5540 5390 & Whatsapp di 0813-1864-0412 dan Email [email protected]